MMTD80 (80 sample thermal demagnetizer). This advanced microprocessor controlled thermal demagnetizer offers the following unique set of features:

  • Simple but sophisticated programmable Heat/Hold/Cool cycle, no need to manually move samples from a heating to a cooling chamber.
  • Internal calibrated solenoid for creating TRM s.  Ideal for ‘double heating’ Thellier experiments.
  • 4 layer Mu-metal shields closed at ends with automatic degaussing and our special electronic activation to improve performance. Samples always in a field <10nT (typically <3nT) during heating and cooling.
  • Any orientation, because the shields are closed at one end the oven it can have a north/south or east west orientation, the field remains low.
  • Fast cycle, 80 samples to 500oC, 25 min hold at 500oC, cool to 40oC in less than 80 minutes, that's an average processing speed of one sample every 60 seconds, fast enough to keep up with any  magnetometer.
  • Non-magnetic and accurate thermocouples, we do not use K type thermocouples as these can become inaccurate with age.
  • Ultra low thermal gradients, Maximum deviation 6oC at 500oC over full load of 80 samples, 4oC over central 40 samples.
  • Absolute temperature accuracy 1oC, resolution 0.1oC, maximum temperature 800oC.

ordering

service

home

 

 

  • Robust ceramic oven element and virtually unbreakable non-magnetic sample holder for long trouble free operation. Built in radiation shields for ultra low thermal gradients.
  • Controlled atmosphere sample holders supplied as standard (vacuum or inert gas).
  • Optional extra MMTDVP 2 stage vacuum pump and fittings.
  • Optional extra MMCCPSU constant current power supply for Thellier experiments, field on during both heating and cooling.
  • 110 or 220V, no additional transformer required. Specify mains voltage when ordering.
  • Battery backup, If mains electricity fails the system remembers the programme sequence and the programme continues once power is restored.
  • Safety Features, no power to oven element when samples are inserted/removed, Software over temperature trip isolates oven and cools, mechanical over temperature trip cuts power to oven. Second 'policeman' controller monitors oven and cuts power if all other systems fail.
  • 3-year parts and labour warranty.

 

 

 

 

 

 

 

 

 